Chapter 13: GAS Monitoring (Mainstream)
To keep secretions and moisture from pooling on the windows, always position the IRMA
probe in a vertical position with the LED pointing upwards.
Warning
Do not place the IRMA airway adapter between the endotracheal tube and an elbow as this
may allow patient secretions to block the adapter windows and result in incorrect
operation.
Warning
Disposable airway adapters shall not be reused. Reuse of the single use adapter can cause
cross infection.
Warning
Used airway adapters shall be disposed of in accordance with local regulations for
biohazardous waste.
Placement of IRMA Probe
When connecting the IRMA probe to an infant patient circuit it is important to avoid a direct
contact between the IRMA probe and the infant's body due to the elevated surface temperature of
the IRMA Probe.
Warning
The IRMA probe is not intended to be in patient contact.
Warning
If, for whatever the reason, the IRMA probe is in direct contact with any parts of the
infant's body an insulation material shall be placed between the IRMA probe and the
body.
Warning
Measurements can be affected by mobile and RF communications equipment. It should be
assured that the IRMA sensor is used in the electromagnetic environment specified in this
manual.
Warning
The IRMA probe is not designed for MRI-environments.
13-8