Chapter 15: BFA Monitoring
When used with electro surgical unit please note the positioning of the neuro sensors. In
order to reduce the hazard of burns, the neuro sensors should not be located between the
surgical site and the electro surgical unit return electrode.
Warning
Not to be used in the presence of flammable gases; explosion risk.
Warning
Pay attention if the BFA monitor is connected to a patient connected to other equipment.
The total of leakage current may exceed the allowable limit and cause a possible hazard to
the patient.
Warning
The conductive parts of neuro sensor should not contact other conductive parts including
earth.
Warning
The monitor should be used in conjunction with other patient monitoring parameters and
clinical signs. This will ensure the optimum balance of the anesthesia/sedation
administration.
Warning
Do not open the BFA case. There are no user-serviceable parts inside. The case should only
be opened by qualified service personnel using proper grounding techniques. When the
case is opened, an electrical shock hazard exists which can result in serious injury to
persons and instrument component damage.
Skin Preparation and Placement of Sensors
To ensure low sensor impedance, clean skin with mild soap and water is recommended as a skin
cleanser.
NOTE:
Alcohol is not recommended as a skin cleanser; it leaves a film layer that may cause
high sensor impedance. If alcohol is used, ensure 30 second dry time.
NOTE:
The performance of the BFA module is only guaranteed by the manufacture when
the BFA Procedure Pack is used.
15-5