Solving Problems; Well-Plate Handler Does Not Turn On When Switched On; The Well-Plate Handler Arm Does Not Move Down Far Enough To Depress The Plate Sensor Switch - Agilent Technologies 1100 Series Reference Manual

Sample capacity extension
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

5
Troubleshooting and Error Handling

Solving Problems

Well-plate Handler does not turn on when switched on.

Possible Cause
Power cord is not firmly plugged in.
Fuse has blown.
Fuse configuration is not correct.
The Well-plate Handler arm does not move down far enough to depress the
plate sensor switch.
Possible Cause
The vertical offset for the position has been
trained too high.
For all other problems, contact Agilent Technologies Product Support.
82
Solution
Verify power cord is firmly seated in receptacles.
Replace fuse(s).
Refer to
"Changing the Fuse Configuration, if
necessary"
on page 18.
Solution
In the software, adjust and save the vertical offset
for the correct plate type.
Note: Place only one well-plate in a rack or device
position when teaching the vertical offset.
1100 Series Sample Capacity Extension

Advertisement

Table of Contents
loading

This manual is also suitable for:

G2257a

Table of Contents