5
Troubleshooting and Error Handling

Solving Problems

Well-plate Handler does not turn on when switched on.

Possible Cause
Power cord is not firmly plugged in.
Fuse has blown.
Fuse configuration is not correct.
The Well-plate Handler arm does not move down far enough to
depress the plate sensor switch.
Possible Cause
The vertical offset for the position has been
trained too high.
For all other problems, contact Agilent Technologies Product Support.
92
Solution
Verify power cord is firmly seated in receptacles.
Replace fuse(s).
Refer to
necessary"
Solution
In the software, adjust and save the vertical
offset for the correct plate type.
Note: Place only one well-plate in a rack or device
position when teaching the vertical offset.
Agilent 1200 Series Sample Capacity Extension User Manual
"Changing the Fuse Configuration, if
on page 20.