Measurements With Thru-Coat D7906 And D7908 Transducers; Enabling The Thru-Coat Function; Figure 5-11 Opening The Thru Coat Setup Dialog Box - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
Press [2nd F], [SAVE/SEND] (WAVEFORM) to save the measured thickness
value and the waveform.
TIP
If you want to always save both the thickness and the waveform when you press
[SAVE/SEND], in the SYSTEM screen, set SAVE DATA to THK+WF (see
"Configuring System Parameters" on page 129).
5.5

Measurements with THRU-COAT D7906 and D7908 Transducers

THRU-COAT is a function that measures the true metal thickness of coated or painted
parts. This function only requires a single back-wall echo and is recommended for
heavy corrosion applications where the outside of the material is coated or painted. If
needed, you can also calibrate the measurement of the coating/paint layer to precisely
measure the thickness of the coating/paint.
5.5.1

Enabling the THRU-COAT Function

The THRU-COAT function is only available when you connect a THRU-COAT
transducer (P/N: D7906 [U8450005] or D7908 [U8450008]) to the 38DL PLUS.
To enable the THRU-COAT function
1.
Connect a THRU-COAT transducer to the 38DL PLUS.
2.
Start the instrument.
3.
Wipe the couplant off the transducer tip.
4.
Press [2nd F], [CAL ZERO] (Do ZERO).
Figure 5‑11 Opening the THRU COAT SETUP dialog box