Figure 5-10 The Active File Name Appearing In The Id Bar - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

current ID. The ID is automatically incremented for the next measurement. When you
press [FILE], the name of the active file appears in the ID bar above the menu (see
Figure 5-10 on page 77).
Figure 5‑10 The active file name appearing in the ID bar
The NONAME00 increment type file, starting with the 001 ID, is the active file by
default when you first use the 38DL PLUS or after resetting the instrument memory.
You can create various types of files and define IDs to represent various 1-D, 2-D, or
3-D thickness measurement locations. When restarting, the instrument automatically
opens the last file that was used.
The following special cases may occur:
When the thickness value is blank, "——-" is saved rather than a value.
When a measurement is already stored at the current ID, the new value
overwrites the old thickness reading unless the overwrite protection is enabled
(see "Setting the ID Overwrite Protection" on page 196).
When the ID increment reaches the end of a sequence and cannot be updated,
Last ID appears on the help text bar, a long beep sound is emitted (when the
beeper is active), and the ID on the display remains unchanged.
Refer to "Using the Datalogger" on page 159 for more information on the datalogger.
To save data at the current ID in the active file
 While the desired thickness value and waveform are displayed, press
[SAVE/SEND] to save the measured thickness value.
OR
DMTA-10004-01EN, Rev. D, November 2016
Name of the active file
Basic Operation
77