3.
In the menu, select ALARM.
4.
In the ALARM screen:
a) Set ALARM ENABLE to ON.
b) Set ALARM MODE to B‑SCAN.
c)
Enter the desired LOW ALARM and HIGH ALARM values.
5.
Press [MEAS] to return to the measurement screen.
6.
The horizontal red alarm lines appear on the B-scan (see Figure 8-5 on page 121).
10.3.3

Saving B-Scans, A-Scans, or Thickness Readings

The 38DL PLUS can do the following tasks while using the B-scan:
Save a live thickness reading while the B-scan is running.
Save any reviewed thickness reading on a frozen B-scan.
Save all thickness readings for one B-scan screen (600 readings) along with the
minimum or maximum thickness readings for a held B-scan.
Save all thickness readings for one B-scan screen (600 readings), the minimum or
maximum thickness reading and waveform for a held B-scan.
Save an entire B-scan history up to 10000 thickness readings, the minimum or
maximum thickness readings and waveform, and the held B-scan.
To save a live thickness reading while the B-scan is running
 Press [SAVE/SEND].
To save any reviewed thickness reading on a frozen B-scan
1.
While the B-scan is running, press [FREEZE] to enter the review mode.
2.
Use the [] and [] keys to review any thickness on the frozen B-scan.
3.
Press [SAVE/SEND] to save to the datalogger, the thickness value for the review
marker position.
To save the minimum or maximum thickness reading for a frozen B-scan
1.
Set B‑SCAN FREEZE MODE to SHOW MINIMUM or SHOW MAXIMUM.
The gage displays the minimum or maximum thickness reading with the
corresponding waveform.
2.
While the B-scan is running, press [FREEZE] to enter the review mode.
DMTA-10004-01EN, Rev. D, November 2016
Using Advanced Gaging Features
145