Olympus 38DL PLUS User Manual page 292

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
Table 38 Typical measurement ranges and default setups for a single element
Setup name
DEFM2-2.25-M207
DEFP2-2.25-M207
DEFM2-20.0-M208
DEFM1-0.5-M101
DEFM1-1.0-M102
DEFM1-1.0-M103
DEFP1-0.5-M2008
a.
The maximum thickness measuring capability depends on transducer type, material condi-
tions, and temperature.
Name
MEAS
OPTION
MEAS TYPE
PROBE TYPE
PULSER
POWER
MAX GAIN
INIT GAIN
TDG SLOPE
MB BLANK
280
Appendix A
transducer
Transducer
M207
Steel: 2.00 mm to 19.00 mm (0.080 in. to 0.750 in.)
M207
Plastic: 2 mm to 12.5 mm (0.080 in. to 0.500 in.)
M208
Steel: 0.50 mm to 10.00 mm (0.020 in. to 0.200 in.)
M101
Steel: 12.5 mm to 635 mm (0.500 in. to 25.00 in.)
M102
Steel: 5.0 mm to 635 mm (0.200 in. to 25.00 in.)
M103
Steel: 2.5 mm to 635 mm (0.100 in. to 25.00 in.)
M2008
Fiberglass: 5.0 mm to 75 mm (0.200 in. to 3.00 in.)
Table 39 Setup parameter description
Description
Echo detect mode
Special measurement modes
Transducer types
Pulser power
Maximum receiver gain
Initial TDG gain
Time Gain slope (default)
Main Bang Blank
(continued)
a
Typical measurement range
Units/Resolutions/Range
Standard dual
Dual echo-to-echo
Thru-Coat
Mode 1
Mode 2
Mode 3
Standard or Oxide (optional)
Barrier Layer (optional)
First Peak
Dual element
Direct contact
Delay line
Immersion
EMAT
60 V, 110 V, 150 V, or 200 V
0.0 dB to 99.8 dB, 0.3 dB steps
0 to MaxGain, 1 dB steps.
0.0 dB/s to 39.9 dB/s
0 ns to 225 µs