DMTA-10004-01EN, Rev. D, November 2016
— When the thickness differential is greater than the % GROWTH value, a high
alarm is indicated.
Table 9 on page 120 shows the result for an example where both % LOSS is set to
20 % and % GROWTH is set to 5 %.

Table 9 Alarm calculation example with the % DIFF mode

Previous
Actual
values
thickness
7.62 mm
6.07 mm
(0.300 in.)
(0.239 in.)
7.62 mm
8.03 mm
(0.300 in.)
(0.316 in.)
7.62 mm
7.24 mm
(0.300 in.)
(0.285 in.)
7.62 mm
7.67 mm
(0.300 in.)
(0.302 in.)
B‑SCAN
The B-scan alarm mode is similar to the standard alarm mode except that in the
B-scan alarm mode, lines are shown in the B-scan grid when the reference values
fall within the B-scan thickness range (see Figure 8-5 on page 121). In addition, the
alarms are operational as you review B-scan thicknesses in B-scan freeze review
mode.
120
Chapter 8
Low alarm
High alarm
Yes
No
No
Yes
No
No
No
No
Calculations
7.62 6.07
-------------------------- -
×
>
100
20%
7.62
or in inches:
0.300 0.239
-------------------------------- -
×
>
100
20%
0.300
8.03 7.62
-------------------------- -
×
>
100
5%
7.62
or in inches:
0.316 0.300
-------------------------------- -
×
>
100
0.300
7.62 7.24
-------------------------- -
×
<
100
20%
7.62
or in inches:
0.300 0.285
-------------------------------- -
×
<
100
20%
0.300
7.67 7.62
-------------------------- -
×
<
100
5%
7.62
or in inches:
0.302 0.300
-------------------------------- -
×
<
100
5%
0.300
5%