Figure 15-2 The Keypad Test Screen - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
2.
In the menu, select TESTS.
3.
In the TESTS screen, select the desired test, and then press [ENTER].
4.
When KEYPAD TEST is selected, in the KEYPAD TEST screen (see Figure 15-2
on page 268):
a) Test any key by pressing it on the keypad.
When that specific key is working correctly, the instrument displays the name
of the key in the last key pressed box.
b) Press [ENTER] to terminate the keypad test.
5.
When VIDEO TEST is selected:
a) In the VIDEO TEST screen, press [DISPLAY] to start the video test.
The display shows three vertical color strips that darken to black. Any
problems with the display show up as a disruption of the pattern.
b) Press [ENTER] to terminate the video test.
6.
When INTERNAL SD CARD TEST or EXTERNAL SD CARD TEST is selected,
in the INTERNAL SD CARD TEST or EXTERNAL SD CARD TEST screen:
a) Press [DISPLAY] to start the test.
The test result appears in SD Card Test Status. The possible results are:
PASS: indicates that the card is functioning properly.
FAIL: indicates that there is a problem with the memory card. When the
external card fails, re-install or replace the card, and then restart the
instrument. When the internal card fails, contact Olympus for service.
268