DMTA-10004-01EN, Rev. D, November 2016
Table 30 General EN15317 specifications (continued)
Parameter
Minimum and maximum
thickness
Parameter
Type
Size
Parameter
Transmitter pulse
Pulser voltage
Pulse rise time
Pulse duration
Parameter
Gain control
Frequency range
Parameter
Data storage
276
Appendix A
Single element: 0.1 mm to 635.0 mm (0.004 in. to 25 in.)
Dual element: 0.5 mm to 635.0 mm (0.020 in. to 25 in.)

Table 31 Display EN15317 specifications

Color graphical TFT, LCD, 640 × 480 pixels
[Height] × [Width], [Diagonal]
56.16 mm × 74.88 mm, 93.6 mm
(2.21 in. × 2.94 in., 3.68 in)

Table 32 Transmitter EN15317 specifications

Adjustable square wave pulser
Pulse voltages: 60 V, 110 V, 150 V, and 200 V
Damping in: 5 ns typical
Damping out: 3.5 ns typical (pulse width dependent)
Adjustable to probe frequency

Table 33 Receiver EN15317 specifications

Automatic or manual: 0 dB to 99 dB
0.5 MHz to 24 MHz typical (filter dependent)

Table 34 Other EN15317 specifications

Internal and external 2 GB microSD memory cards.
Each card:
475000 thickness readings or
20000 waveforms with thickness readings
Value