Figure 8-5 Example Of A B-Scan Alarm Mode - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

NOTE
The thickness value and the alarm indicator appear in color only when the indoor
color scheme is active (see "Color Schemes" on page 52 to change the color scheme).
REDUCTION RT
The REDUCTION RT option appears only when the active file is configured with
FILE DATA MODE set to REDUCTION RT. You set the reduction percentage for
a low (YELLOW ALARM) and high (RED ALARM) limits, and the instrument
displays (see Figure 8-6 on page 122):
— The RED indicator for reduction rates that are greater than or equal to the
RED ALARM value.
— The YEL indicator for reduction rates that are between the YELLOW ALARM
and the RED ALARM values.
— The GRN indicator for reduction rates that are below the YELLOW ALARM
value.
Figure 8‑5 Example of a B‑scan alarm mode
DMTA-10004-01EN, Rev. D, November 2016
Red alarm limits
Using Special Functions
121