Using The B-Scan; Using The B-Scan Alarm Mode - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
10.3.1

Using the B-Scan

The following procedure describes how to activate and use the B-scan.
To use the B-scan
1.
While in the measurement screen, press [SETUP MENU].
2.
In the menu, select B-SCAN.
3.
In the B-SCAN screen, set B‑SCAN ENABLE to ON, and configure the other
B-scan parameter as desired (see "B-Scan" on page 139).
4.
Press [MEAS] to return to measurement screen with the B-scan active.
The B-scan begins drawing the cross section of the material as soon as the first
non-LOS reading is collected.
5.
To stop the collection of B-scan data:
Press [FREEZE].
OR
With B‑SCAN LOS MODE set to STOP ON LOS, simply uncouple the
transducer from the test material.
6.
While the B-scan is frozen, if B‑SCAN FREEZE REVIEW is set to ON, use the []
and [] keys to move the review marker and read the thickness value
corresponding to the marker position.
7.
Press [MEAS] again to reset the B-scan and begin a new B-scan.
8.
Refer to "Saving B-Scans, A-Scans, or Thickness Readings" on page 145 for details
on how to save B-scan data.
10.3.2

Using the B-Scan Alarm Mode

You can specify B-scan low and high alarm reference values and turn the visual and
audible alarm functions ON and OFF. The B-scan alarm mode is similar to the
standard alarm mode (see "Using Alarms" on page 117) except that the alarm lines
are shown in the B-scan grid, as long as the alarm reference values fall within the
B-scan thickness range. The alarms are also operational as you review B-scan
thicknesses in review mode, in a frozen B-scan.
To use the B-scan alarm mode
1.
Activate and configure the B-scan (see "Using the B-Scan" on page 144).
2.
While in the measurement screen, press [SETUP MENU].
144
Chapter 10

Advertisement

Table of Contents
loading