Figure 13-1 The Active Screen For The Setup Of A Single Element Transducer - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
To create a custom setup for a single element transducer
1.
Connect the single element transducer to the 38DL PLUS (see "Setting Up the
Transducer" on page 63).
2.
Press [XDCR RECALL].
3.
In the menu, select the CUSTOM SINGLE ELEMENT.
4.
In the CUSTOM SINGLE ELEMENT screen, select the desired custom setup
location (SE‑USER‑n) in which you wish to save the custom setup.
TIP
To reduce the number of changes to make to parameter values, you can also select an
existing single element setup with parameter values that are close to the desired
setup.
5.
In the ACTIVE screen (see Figure 13-1 on page 222):
a) Set DET MODE to the desired detection mode (see "Detection Modes" on
page 225 for details).
Figure 13‑1 The ACTIVE screen for the setup of a single element transducer
b) Enter a SETUP NAME that describes the transducer and/or the application
for which you create the setup.
c)
Set MEAS TYPE to the desired measurement type. The available choices are:
222
Chapter 13