Table 17 Resulting Id Example For The Boiler File Type - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

Table 17 Resulting ID example for the BOILER file type

ELEVATIONS
10FT
20FT
45FT
100FT
To create a boiler data file
1.
From the measurement screen, press [FILE], and then select CREATE in the menu
(see "Creating a Data File" on page 163 for details on the first parameters).
2.
At the bottom of the CREATE screen, select CONTINUE.
3.
In the second page of the CREATE screen (see Figure 11-13 on page 180):
a) Enter the START TUBE and END TUBE values.
b) Enter two or more CUSTOM POINTS values.
c)
Press [2nd F], [] to finish entering CUSTOM POINTS values.
d) Enter two or more ELEVATIONS values.
e)
Press [2nd F], [] to finish entering ELEVATIONS values.
f)
In INC 1ST BY, select which parameter increments first (POINT, TUBE, or
ELEVATIONS).
g) In INC 2ND BY, select which parameter increments second (POINT, TUBE,
or ELEVATIONS).
h) Select CREATE.
START TUBE
END TUBE
01
73
DMTA-10004-01EN, Rev. D, November 2016
CUSTOM
Resulting IDs
POINTS
L (left)
10FT-01L
C (center)
10FT-01C
R (right)
10FT-01R
10FT-02L
...
10FT-73R
20FT-01L
...
100FT-73R
Using the Datalogger
179

Advertisement

Table of Contents
loading

Table of Contents