14.2 Setting Up USB Communication

The default communication protocol for the 38DL PLUS is USB 2.0. The 38DL PLUS
can also be set to use RS-232 serial communication (see "Setting Up RS-232 Serial
Communication" on page 245).
To set up the USB communication
1.
Ensure that the 38DL PLUS driver is installed on the computer.
This driver is installed when you install the GageView interface program.
NOTE
For more information on installing GageView, refer to the GageView Interface Program
— User's Manual (P/N: 910-259-EN [U8778347]).
2.
If you use a communication program other than GageView, refer to the program
documentation to properly configure the program for USB communication.
3.
Turn the 38DL PLUS on.
4.
From the measurement screen, press [SETUP MENU].
5.
In the menu, select COMM.
6.
In COMM screen, set CONNECTION TYPE to USB.
7.
Press [MEAS] to return to the measurement screen.
8.
Connect one end of a USB cable to the USB client connector under the I/O door of
the 38DL PLUS, and the other end to a USB port of the computer (see Figure 14-1
on page 245).