Saving Data; Figure 5-9 Coupling Dual Element Transducer And Reading The Measured Thickness - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
To measure thicknesses
1.
Apply couplant to the test block or tested part at the measuring spot.
NOTE
In general, use a thinner couplant such as propylene glycol, glycerin, or water for
smooth material surfaces. Rough surfaces require a more viscous couplant such as gel
or grease. Special couplants are required for high temperature applications.
2.
Using moderate to firm pressure, couple the tip of the transducer to the surface of
the test material, and keep the transducer as flat as possible on the material
surface (see Figure 5-9 on page 76).
Figure 5‑9 Coupling dual element transducer and reading the measured thickness
3.
Read the measured thickness value for the tested part.
5.4

Saving Data

The 38DL PLUS datalogger is a file based system where one file is opened at a time.
The active file stores a measurement at a thickness measurement location ID. Each
time you press [SAVE/SEND], the displayed value is saved to the active file at the
76
Chapter 5
12.7 mm
(0.500 in.)