Olympus 38DL PLUS User Manual page 124

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
3.
Press [XDCR RECALL].
4.
In the menu, select DEFAULT HP SINGLE ELEMENT.
5.
In the DEFAULT HP SINGLE ELEMENT screen, highlight the default transducer
setup for the M2008 (DEFP1‑0.5‑M2008) or any custom setup that uses the M2008
transducer.
6.
Press [MEAS] to return to the measurement screen with the recalled transducer
setup.
7.
Wipe couplant off the transducer tip.
8.
Press [2nd F], [CAL ZERO] (Do ZERO).
9.
Perform the material sound velocity and zero calibration (see "Calibrating the
Instrument" on page 67).
112
Chapter 7