Figure 11-22 Warning Message When Resetting Measurements - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

Using the measurement reset erases all files and data contained in those files. The
deleted files and the data they contain cannot be recovered. The datalogger will be
completely empty after this procedure.
To delete all files
1.
Press [2nd F], [SETUP MENU] (SP MENU).
2.
In the menu, select RESETS.
3.
In the RESETS screen (see Figure 11-22 on page 191):
a) In the RESETS list, select INTERNAL MEMORY RESET to delete all files on
the internal microSD memory card.
b) Select RESET to delete all the files.
OR
Select CANCEL or press [MEAS] to abort the operation.
Figure 11‑22 Warning message when resetting measurements
CAUTION
DMTA-10004-01EN, Rev. D, November 2016
Using the Datalogger
191