Figure 10-17 Examples Of The Measurement Screen With The Avg/Min Measurement Active - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

A indicating active Average
Figure 10‑17 Examples of the measurement screen with the AVG/MIN
To make thickness readings using the AVG/MIN measurement
1.
Activate and configure the AVG/MIN function (see "Configuring AVG/MIN
Measurements" on page 153).
2.
Couple your transducer to the material.
The thickness appears on the display.
3.
Press [SAVE/SEND] to display the thickness in the left most thickness box.
Each time you press [SAVE/SEND] the actual thickness reading appears in the
next shaded box moving to the right while the minimum or average of the
thickness readings is automatically calculated and placed in the right most box.
4.
If you want to replace a reading, use the [] and [] keys to switch from one
highlighted box to another, and press [SAVE/SEND] to save the new reading.
5.
When the minimum or average box is highlighted, press [SAVE/SEND] a last
time to store that value in the datalogger at the current ID# location, and clear all
the boxes.
OR
Press [MEAS] to clear all the boxes without saving to the datalogger.
DMTA-10004-01EN, Rev. D, November 2016
M indicating active Minimum
measurement active