7.3.2

Setting Up for Oxide Layer Measurement

You need to connect an appropriate transducer, select the corresponding default
setup, and configure the oxide and material measurement parameters.
To set up for oxide layer measurement
1.
Connect the M2017 or M2091 transducer to an LCM-74-4 cable.
2.
Connect the cable to the T/R 1 single-element transducer connector at the top of
the 38DL PLUS.
3.
Press [ON/OFF] to start the instrument.
4.
While in the measurement screen, press [XDCR RECALL].
5.
In the DEFAULT SINGLE ELEMENT screen, highlight DEF‑OXIDE/M2017 or
DEF‑OXIDE/M2091 depending on which transducer you are using.
6.
Press [MEAS].
7.
Press [SETUP MENU].
8.
In the menu, select OXIDE.
9.
In the OXIDE screen (see Figure 7-2 on page 95):
a) In OXIDE MEAS TYPE, select the type of measurement for the oxide layer
(THICKNESS or TIME OF FLIGHT).
b) In MATERIAL MEAS TYPE, select the type of measurement for the material
layer.
Figure 7‑2 The OXIDE screen
DMTA-10004-01EN, Rev. D, November 2016
Software Options
95