When DIFF MODE is set to REDUCTION RT only:
d) In FORMER THICKNESS, enter the original thickness value, as measured
before bending the metal.
e)
In LARGE FONT, select which measurement appears at the bottom of the
measurement screen with the large font (THICKNESS or REDUCTION RT).
4.
Press [MEAS] to return to the measurement screen with the displayed differential
value.
8.2
Using the Minimum, Maximum, or Min/Max Thickness Mode
You can activate the minimum, maximum, or min/max thickness mode to also display
retained minimum and/or maximum thickness values. The MIN and/or MAX values
appear on the left side of the main thickness reading (see Figure 8-3 on page 116). The
minimum or maximum thickness value replaces the main thickness display when the
transducer is uncoupled or when a loss-of-signal (LOS) occurs. The replacing value
appears with outlined digits.
Figure 8‑2 The DIFF screen
DMTA-10004-01EN, Rev. D, November 2016
Using Special Functions
115