Figure 12-4 Answering Yes At The V-Path Calibration Prompt - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
NOTE
The thickness readings at this point may not be accurate because a proper zero offset
for the transducer has not yet been set.
c)
Couple the transducer to the thinnest test block step.
d) If necessary, press [GAIN] to adjust the gain so that the instrument properly
detects the back-wall echo.
e)
If necessary, adjust the extend blank so that the gage is making proper
detection on the back-wall signal (see "Blanking Adjustments in Manual
Echo-to-Echo Detection Mode" on page 83).
f)
Ensure that the instrument properly detects echoes on all test block steps.
4.
Activate the V-path function (see "Activating the V-Path Function" on page 216
for details).
5.
Couple the transducer to a thick sample representing the thickest material that
you will measure.
6.
Press [CAL VEL].
7.
In the help text bar, select YES to the VPath Calibration? prompt (see Figure 12-4
on page 218).
Figure 12‑4 Answering YES at the V‑path calibration prompt
8.
While the transducer is coupled to the thick sample and getting a steady thickness
reading, press [ENTER].
9.
In the V‑PATH CALIBRATION screen, edit the point 1 value to match the known
thickness (see Figure 12-5 on page 219).