Figure 5-5 Performing The Zero Calibration On A 5-Step Test Block - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

DMTA-10004-01EN, Rev. D, November 2016
2.54 mm
(0.100 in.)
Figure 5‑5 Performing the zero calibration on a 5‑step test block
3.
Press [MEAS] to complete the calibration and return to the measurement screen.
IMPORTANT
If you turn off the instrument before pressing [MEAS], the velocity is not updated to
the new value; instead the instrument retains the previous value.
NOTE
When the 38DL PLUS detects an error in the calibration procedure, it successively
displays the following messages in the help text bar before returning to the
measurement screen:
"Potential wrong echo detected!"
"Invalid calibration results!"
In this case, the velocity is not changed. The probable case is that an incorrect
thickness value was entered.
Basic Operation
69