Gage hanging up on trailing edge of back wall 1
Because the mode 3 echo blank limits minimum measurable thicknesses, set the
controls as short as possible; typically no longer than a few hundred nanoseconds.
Special cases may be an exception where mode conversion effects on curved samples
cause significant spurious signals between legitimate echo peaks; in such cases set the
mode 3 echo blank as long as necessary to prevent detection of the spurious signals.

13.9 Saving Setup Parameters

After adjusting selected waveform parameters, you can store the settings for a quick
and easy recall. The 38DL PLUS can store up to 35 custom setups in its internal
memory.
To save setup parameters
1.
Make the appropriate changes to the waveform parameters.
2.
Press [XDCR RECALL].
In the menu, ensure that ACTIVE is highlighted. The changes that you made are
located in the active setup.
3.
Press [SAVE/SEND].
4.
In the SAVE SETUP screen (see Figure 13-17 on page 240):
a) In SAVE AS, enter a name for the custom setup.
Figure 13‑16 Examples of M3 Blank adjustment
Custom Setups for Single Element Transducers
DMTA-10004-01EN, Rev. D, November 2016
M3 BLANK set properly