6.1

Connecting the E110-SB EMAT Transducer

You can use the E110-SB [U8471001] EMAT transducer with the 38DL PLUS. You need
to use the EMAT 1/2XA/E110 [U8767104] filter adaptor and a standard LEMO-to-BNC
cable (P/N: LCB-74-4 [U8800320]) as shown in Figure 6-1 on page 88.
(P/N: 1/2XA/E110 [U8767104])
Figure 6‑1 EMAT transducer and filter adaptor connection
The 38DL PLUS automatically detects the E110-SB and 1/2XA/E110 filter adaptor and
automatically recalls the default DEF‑EMAT/E110 setup from the DEFAULT SINGLE
ELEMENT transducer list. Like for standard transducers, you may need to adjust the
gain and the extended blank to make proper thickness measurements. Experienced
technicians familiar with the basic theory of ultrasonic gaging and the interpretation
of ultrasonic waveforms can also press [WAVE ADJ] to adjust other transducer
parameters.