DMTA-10004-01EN, Rev. D, November 2016
3.
Set BEEPER to ON or OFF (see "Setting the User Interface Language and Other
System Options" on page 49 for details).
4.
Set INACTIVE TIME to ON or OFF (see "Setting the User Interface Language
and Other System Options" on page 49 for details).
5.
In LANGUAGE, select the desired user interface language (see "Setting the User
Interface Language and Other System Options" on page 49 for details).
6.
Set RADIX TYPE to the desired character (period or comma) to separate the
integer and the decimal numbers.
7.
Set SAVE/SEND KEY to configure what happens to the actual data (as set in
SAVE DATA) when you press [SAVE/SEND]. The options are:
— SAVE: saves the actual data to the internal datalogger.
— SEND: sends the actual data to the RS-232 port (see "Setting Up RS-232 Serial
Communication" on page 245 for details on RS-232 data communication).
NOTE
The SAVE/SEND KEY parameter also affects the save/send operation triggered using
the optional foot switch. The 38DL PLUS cannot send a single thickness value through
the USB communication port.
8.
Set SAVE DATA to save only the thickness measurement (THICKNESS) or both
the thickness and the waveform (THK+WF).
130
Chapter 9
Figure 9‑2 The SYSTEM screen