Table 27 Measurement Default Settings - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

Parameter
Measurement mode with differential
Material sound velocity
Differential reference value
Low alarm reference value
High alarm reference value
Display update rate
Zoom
Range
LOS condition
Resolution
INTERNAL MEMORY RESET
Deletes all stored data on the internal microSD memory card and reformats the
card.
CAUTION
The internal memory reset permanently deletes all stored thickness
readings/waveforms that are stored in the memory card. However, this reset does not
delete any stored transducer setups.
CLOCK RESET
Resets the date to 01/01/2010 in the MM/DD/YYYY format and the time to
12:00AM in the 12-hour format.
COMMUNICATION RESET
Changes the communication settings to the default factory values listed in
Table 28 on page 266.

Table 27 Measurement default settings

Maintaining and Troubleshooting the 38DL PLUS
DMTA-10004-01EN, Rev. D, November 2016
Default value
Fast, min, max, and alarms turned off
5.969 mm/s or 0.2350 in./s
(approximate velocity of the included test
blocks)
0.0 mm or 0.0 in.
0.0 mm or 0.0 in.
635.0 mm or 25.0 in.
4 per second
Off
Minimum range
Blank thickness display
0.01 mm or 0.001 in.
265