Figure 12-1 The Generic Setup Selection Screen; Table 21 Default Transducers - Olympus 38DL PLUS User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 38DL PLUS:
Table of Contents

Advertisement

Figure 12‑1 The GENERIC SETUP SELECTION screen
3.
Press [MEAS] to return to the measurement screen.
4.
Press [XDCR RECALL].
5.
In the menu, select DEFAULT DUAL ELEMENT.
6.
In the DEFAULT DUAL ELEMENT screen, among the default transducer list,
select the transducer that best matches the frequency and the diameter of the dual
element transducer that you are using (see Table 21 on page 213).
Default transducer
D790/791/D799
D792/D793
D794/D795
D797
D798/D7226
D7912/D7913
MTD705
D7906 (THRU-COAT)
D7908
Generic setups for dual element transducers:
DEF-DE-2MHZ

Table 21 Default transducers

Frequency
5 MHz
5 MHz
10 MHz
2 MHz
7.5 MHz
10 MHz
5 MHz
5 MHz
7.5 MHz
2 MHz
Dual Element Transducer Setups
DMTA-10004-01EN, Rev. D, November 2016
Tip diameter
11.0 mm (0.434 in.)
7.2 mm (0.283 in.)
7.2 mm (0.283 in.)
22.9 mm (0.900 in.)
7.2 mm (0.283 in.)
7.5 mm (0.295 in.)
5.1 mm (0.200 in.)
11.0 mm (0.434 in.)
7.2 mm (0.283 in.)
213