In mode 2 and mode 3, the echo window is limited to the time interval between
successive interface echoes (see Figure 13-12 on page 234). The end of the echo
window must be set ahead of the second interface echo to prevent an incorrect
detection, which in turn determines a maximum measurable thickness. In
applications involving mode 2 and mode 3 immersion measurements, the echo
window placement must apply throughout the range of the water paths to be used.
Interface echo
Interface blank
Figure 13‑12 Echo window setting for mode 2 and mode 3
234
Chapter 13
Figure 13‑11 Echo window setting for mode 1
Echo window
Echo window
Second interface echo