New Folder

Reset/Recall/Power-On State

Model N6705 User's Guide
To create a new folder at the present directory level, press the File
key, then scroll to and select File Management. In the Action
dropdown box, select New Folder.
Parameter:
Description:
Path\New
Specifies a name for the folder.
Folder Name
Internal:\ specifies the instrument's internal memory.
External:\ specifies the Memory port on the front panel.
Enter the name in the text field. See "Enter the File Name"
Browse
Lets you browse another directory or USB memory device.
Create Folder
Creates the new folder in the specified location.
As shipped, the DC Power Analyzer is configured to automatically
recall the Reset State (*RST) settings at power-on. However, you can
configure the reset, recall, and power-on state of the instrument.
Press the File key, then scroll and select Reset/Recall/Power-On
State.
Selecting Reset to Defaults lets you immediately return the
instrument to its factory default settings as described in chapter 1.
Quick Save/Recall lets you save and subsequently recall an
instrument state in memory locations 0 through 9. This is the same
as saving an instrument state to a file name, but quicker. These
functions can also be accessed using the SCPI *SAV and *RCL
commands.
At Power-On lets you recall the Reset State (*RST), or recall the
instrument state stored in location 0.
Using the System Functions 5
143