Calibration/Adjustment

Available Features
Calibration/adjustment can only be performed when
– there is no load on the balance/scale,
– the balance/scale is tared,
– the internal signal is stable.
If these conditions are not met, an error message is displayed. Otherwise, the weight
required for external calibration/adjustment is displayed.
Calibration/adjustment of verified balances/scales of accuracy class K or l :
– Calibration/adjustment is blocked (protective cap over menu access switch is sealed)
> Calibration/adjustment is possible only after the seal has been removed. In this case,
the verification becomes null and void and the balance/scale must be re-verified.
Example
Step
1. To unblock calibration/adjustment
function, open the cover plate
on the bottom of the balance/scale
and move the menu access switch
to the "open" position
2. Tare the balance/scale
3. Begin calibration
Calibration weight is displayed
without weight unit (here: 1000 g)
4. Place the indicated calibration
weight on the balance/scale
After calibration, the calibration
weight is displayed with wt. unit
5. Remove the calibration weight
8
Key (or instruction) Display/Printout
t
+
t >2 sec.
+
0.0
g
1000.0
1000.0
1000.0
g
0.0
g