U2000 Series Power Sensors Specification - Keysight N1913B EPM Series Service Manual

Power meters
Table of Contents

Advertisement

[a] Within one hour after zero set, at a constant temperature after a 24-hour warm-up of the power meter.
[b] The number of averages at 16 (for normal mode) and 32 (for x2 mode), at a constant temperature, measured over a
[c] Specification applies to the low power path, up to 75% relative humidity.
[d] The zero drift and measurement noise specifications are tested at 50 MHz.
Table 1-5
Power sensor
U2000/1/2A
U2004A
U2000/1/2H
Keysight N1913B/N1914B EPM Series Power Meters Service Guide
one minute interval and two standard deviations. For Keysight E-Series power sensors, the measurement noise is
measured within the low range. Refer to the relevant power sensor manual for further information.

U2000 Series power sensors specification

–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–50 dBm to –25 dBm
–28 dBm to –5 dBm
Specifications and Characteristics
Range
–7 dBm to 15 dBm
10 dBm to 20 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
–10 dBm to 1 dBm
–1 dBm to 5 dBm
3 dBm to 25 dBm
20 dBm to 30 dBm
[a]
Zero drift
Measurement
200 pW
400 pW
1.5 nW
50 nW
500 nW
2 µW
200 pW
400 pW
1.5 nW
50 nW
500 nW
2 µW
2 nW
4 nW
15 nW
500 nW
5 µW
20 µW
1
[b]
noise
1 nW
1.5 nW
15 nW
650 nW
1 µW
10 µW
1 nW
1.5 nW
15 nW
650 nW
1 µW
10 µW
10 nW
15 nW
150 nW
6.5 µW
10 µW
100 µW
33