2. Press DC Bias > DC Bias State and toggle between the ON|OFF
selection to turn ON/OFF the dc bias state.
bar in the lower right area of the screen.
Optimizing dc bias range
1. Press
Sweep
2. Press DC Bias > Meas Range and select the desired the dc bias range
to within 1mA|10mA|100mA selection options.
DC bias delay
1. Press
Sweep
2. Press DC Bias > DC Bias Delay and input the delay time value.

Setting Trigger

Overview
Selecting a Method to Start Measurement (Trigger Source)
Selecting Sweep Trigger/Measurement Point Trigger
Setting the External Trigger
Controlling Trigger
Sweeping Multiple Sweep Ranges with Different Conditions in a
Single Action (Segment Sweep)
Other topics about Setting Measurement Conditions
Overview
The E4990A has one trigger source. When this trigger source detects a
trigger signal that has occurred, a sweep or point measurement is
performed for channels.
The execution of measurement for each channel does not depend on
whether the channel is displayed. Channels that have been activated can
be measured even if they are not displayed.
For each channel, a sweep is performed only for the stimulus ports
required to update the parameters of the displayed trace.
Selecting a Method to Start Measurement (Trigger Source)
The trigger source generates a cue signal that initiates a measurement
process. Four types of trigger sources are available:
1. Press
Trigger
92
Turning dc bias ON displays
Setup.
Setup.
> Trigger Source, then select the desired trigger source.
at the instrument
DC ON

Hide quick links: