Setting The Cx3300 For An Edge Trigger - Keysight CX3300 Series User Manual

Device current waveform analyzer
Table of Contents

Advertisement

Using the CX3300
Setting Up Triggers

Setting the CX3300 for an Edge Trigger

1. Press the
2. Press the
3. Press the
4. Turn the Level knob to adjust the current or voltage level at which the
58
key until the desired source LED is lit.
Source
You can choose any of the channels or the Aux Trig or Line input as the source
for an edge trigger.
key until the desired slope LED is lit.
Slope
You can have an edge trigger on a rising or falling edge, or both.
key until the desired LED is lit (Trig'd or Auto).
Sweep
When Trig'd is selected, the CX3300 must find the trigger before capturing
and displaying data.
When Auto is selected, if a trigger does not occur within a certain amount of
time, an acquisition is automatically saved and displayed. In Auto trigger
mode, you are able to see your signals while setting up the desired trigger.
CX3300 will trigger.
You can also drag the trigger level indicator at the left side of the display to
adjust the level. See
Pressing the knob changes the trigger level to half.
Figure 2-4
for the trigger level indicator.
Keysight CX3300 User's Guide, Edition 4