Settings Conflict; Param Only Available When Frequency Mode Is Swept - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict; Param
only available when
Frequency Mode is Swept
-221
Settings conflict; Param
only available when valid
cal data exists
-221
Settings conflict; PAvT
step end time exceeds the
analysis interval
-221
Settings conflict; PAvT
step start time is less
than zero
-221
Settings conflict; Phase
Discontinuity measurement
cannot be averaged, Acq=#
Step=#
-221
Settings conflict; Power
search is not available
when ALC is On
-221
Settings conflict; Power
search is not available
when RF is Off
-221
Settings conflict; Power
search is not available
when trigger type is not
free-run
-221
Settings conflict; Preamp
gain is not available in
this Mode
-221
Settings conflict; Preamp
unavailable with
electronic attenuator on
-221
Settings conflict; Presel
Center not available for
unpreselected external
mixers
-221
Settings conflict;
Preselector Adjust not
available for
unpreselected Ext mixer
Instrument Messages
Verbose/Correction Information
SCPI only message. This parameter is only available when the
Freq mode is set to Swept. Change the Freq Mode to Swept
SCPI only message. The 'Apply Calibration' parameter is only
available when the stored cal data matches the current setup.
Perform a fresh 'Calibrate Now' or change setup such that current
cal data is valid.
The end time of PAvT measurement exceeds the analysis interval.
Decrease the end time of PAvT or increase the Analysis interval.
The start time of PAvT measurement cannot be less than ZERO.
The average count of acquisition which contains Phase
Discontinuity measurement cannot be greater than 1.
Set the average count of current acquisition to 1. Or remove the
Phase Discontinuity measurement from this analysis step.
Cannot do power search when ALC is On.
Cannot do power search when RF is Off.
Cannot do power search when trigger type is not free-run.
Preamp gain correction is not available in some Modes or
Measurements
The electronic attenuator is on. Internal preamp cannot be used
while you are using the electronic attenuator.
The selected mixer type is Unpreselected. Preselector power
centering cannot be done on an unpreselected mixer.
The selected mixer type is Unpreselected. The preselector
adjustment cannot be done on an unpreselected mixer.
61