Measurement Accuracy; Amplitude Repeatability - Keysight M9290A Manual

Cxa-m pxie signal analyzer
Table of Contents

Advertisement

Description

Measurement Accuracy

Phase Noise Density Accuracy
c
Default settings
Overdrive On setting
RMS Markers
a. This does not include the effect of system noise floor. This error is a function of the signal (phase noise
of the DUT) to noise (analyzer noise floor due to phase noise and thermal noise) ratio, SN, in decibels.
The function is: error = 10 × log(1 + 10
For example, if the phase noise being measured is 10 dB above the measurement floor, the error due to
adding the analyzer's noise to the UUT is 0.41 dB.
b. Offset frequency errors also add amplitude errors. See the Offset frequency section, below.
c. The phase noise density accuracy is derived from warranted analyzer specifications. It applies with
default settings and a 0 dBm carrier at 1 GHz. Most notable about the default settings is that the Over-
drive (in the advanced menu of the Meas Setup menu) is set to Off.
d. The accuracy of an RMS marker such as "RMS degrees" is a fraction of the readout. That fraction, in per-
cent, depends on the phase noise accuracy, in dB, and is given by 100 × (10
1). For example, with +0.30 dB phase noise accuracy, and with a marker reading out 10 degrees RMS, the
accuracy of the marker would be +3.5% of 10 degrees, or +0.35 degrees.
Description

Amplitude Repeatability

(No Smoothing, all offsets, default
settings, including average = 10)
a. Standard deviation. The repeatability can be improved with the use of smoothing and increasing
number of averages.
Specifications
ab
±1.18 dB
SN/10
)
Specifications
Supplemental Information
±1.06 dB (nominal)
d
See equation
PhaseNoiseDensityAccuracy / 20
Supplemental Information
a
< 1 dB (nominal)