Power Measurement Accuracy Test - Keysight E5052B Service Manual

Signal source analyzer
Hide thumbs Also See for E5052B:
Table of Contents

Advertisement

3. Power Measurement Accuracy Test

Description
This test checks the RF power measurement accuracy of the E5052B. The measurement
power level for this test is set at -20 dBm to 20 dBm in 10 dB step.
Test equipment
Signal Generator
RF Amplifier (AMP 1)
RF Amplifier (AMP 2)
Power Meter
Power Sensor
Step Attenuator
Power Splitter
Blocking Capacitor
N(f)-SMA(m) adapter, 2 ea.
3 dB Fixed Attenuator, 2 ea.
N(m)-N(m) cable 61cm, 3 ea.
N(m)-N(m) Adapter
BNC cable, 122cm
Chapter 2
Individual Test Description
Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
R&K A3000-2H-R
Keysight 83020A
Keysight E4419B
Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Keysight 8496H
Keysight 11667A
Keysight 11742A
Keysight p/n 1250-1562
Keysight 8491A Opt.003
Keysight N6314A (p/n 8120-8862)
Keysight p/n 1250-1475
Keysight p/n 8120-1840
Performance Tests
21

Advertisement

Table of Contents
loading

Table of Contents