Application examples
In the examples that follow, substitute your own application details. If the examples are
not suitable to your application, check the relevant parameter references for other
available options.
Level application
Distance
8.5 m
Quick Start Parameter
Transducer
Operation
Temperature Source
Units
Empty
Span
Response Rate
1.
2.3.2. Fill Rate/minute on page 124
See
The application is a vessel that takes an average 3 hours (180 minutes) to fill and 3 weeks
to empty.
Fill rate = (
Empty
= (15.5 m – 1 m) / 180 min.
= 14.5 m /180 min. = 0.08 m/min.
Page 58
HydroRanger 200 HMI – OPERATING INSTRUCTIONS
1.0 m
Space
7.5 m
Level
6.5 m
Setting
XPS-15
Level
TS-3
m
15.0
1.0
Slow
–
) / fastest of fill or empty time
Span
Sensor reference point
Span
15.0 m
Empty
Range Extension
Description
Transducer to be used with the HydroRanger.
Material level referenced from Low Cal. Point.
Temperature source.
Sensor measurement units.
Process empty level.
Process full level.
1
Sets Fill Rate
/ Empty Rate to 0. 1 m/minute.
.
A5E36281317