Calibrating 4 mA Output
1.
Connect the mA receiving device to the HydroRanger 200 HMI.
2.
Put the HydroRanger 200 HMI into Navigation View.
2.5.9. Milliamp Output
3.
Set
4.
View the mA level on the receiving device.
5.
If there is a discrepancy, do the following steps:
a. Attach an ammeter to HydroRanger 200 HMI mA output.
b. Enter the exact value displayed on the ammeter into
c. The ammeter should then read exactly 4.00 mA.
The unit is now calibrated for 4 mA for the receiving device.
Calibrating 20 mA Output
1.
Connect the mA receiving device to the HydroRanger 200 HMI.
2.
Put the HydroRanger 200 HMI into Navigation View.
2.5.9. Milliamp Output
3.
Set
4.
View the mA level on the receiving device.
5.
If there is a discrepancy, do the following steps:
a. Attach an ammeter to HydroRanger 200 HMI mA output.
b. Enter the exact value displayed on the ammeter into
c. The ammeter should then read exactly 20.00 mA.
The unit is now calibrated for 20 mA for the receiving device.
A5E36281317
to 4.0 for the selected mA output.
to 20.0 for the selected mA output.
HydroRanger 200 HMI – OPERATING INSTRUCTIONS
2.5.11. 4 mA Output Trim
2.5.12. 20 mA Output Trim