2.11.2.10. Short Shot Bias
2.11.2.11. Short Shot Floor
2.11.2.12. Short Shot Range
A5E36281317
Enter the value (in percent of echo height) to ensure the Echo Lock
Window intersects the Echo Profile at the sharpest rising portion of
the Echo Profile representing the true echo.
Slants the echo evaluation in favor of the short shot echo when both
short and long shots are evaluated (see 2.1.17. Shot/pulse Mode on
page 127).
Single-point model
Index
Global
Range: 0 to 100
Values
Preset: 20
Related
Enter the minimum echo strength, in dB above 1 uV, derived from a
short shot to be considered for evaluation.
Single-point model
Index
Global
Range: 0 to 100
Values
Preset: 50
Related
Enter the maximum distance in 2. 1 .1. Units to be measured using short
shot echoes.
Single-point model
Index
Global
Range: 0 to 10 m or equivalent ,depending on
Values
Preset: 1.000
2. 1 .5. Transducer
Altered by
Related
This feature is automatically altered when
HydroRanger 200 HMI – OPERATING INSTRUCTIONS
2. 1 . 1 7. Shot/pulse Mode
2. 1 . 1 1. Number of Short Shots
2. 1 .7. Short Shot Frequency
2. 1 . 1 0. Long Shot Duration
2. 1 1.2. 1 1. Short Shot Floor
2. 1 1.2. 1 2. Short Shot Range
2. 1 . 1 1. Number of Short Shots
2. 1 .7. Short Shot Frequency
2. 1 . 1 0. Long Shot Duration
2. 1 1.2. 1 0. Short Shot Bias
2. 1 1.2. 1 2. Short Shot Range
2. 1 . 1 1. Number of Short Shots
2. 1 .7. Short Shot Frequency
2. 1 . 1 0. Long Shot Duration
2. 1 1.2. 1 0. Short Shot Bias
2. 1 1.2. 1 1. Short Shot Floor
2. 1 .5. Transducer
Dual-point model
Transducer
Dual-point model
Transducer
Dual-point model
Transducer
2. 1 . 1 . Units
is altered.