Chapter 7 Verifying the Instrument Performance

Creating a Plate Document for the RNase P Verification Run

Creating a Plate Document for the RNase P Verification Run
In the SDS software, click
1.
select File New).
In the New Document dialog box:
2.
Select Assay Absolute Quantification
a.
(Standard Curve).
Select Container 96-Well Clear.
b.
Select Template AQ RNase P Install.
c.
In the Operator field, enter your name.
d.
In the Comments field, enter any
e.
information that you want to attach to the
file (such as the plate bar code).
In the Default Plate Name field, enter the
f.
bar code of the RNase P plate.
Click
.
g.
In the SDS software, select File Save As.
3.
Do not modify the plate
IMPORTANT!
document. The sample, detector, and method for
the run are coded into the software.
Save the document. In the Save As dialog box,
4.
select the folder:
If the Save in field does not display SDS
Documents, navigate to D: drive Applied
Biosystems SDS Documents.
After selecting the folder, Click
Prepare and run the RNase P plate as explained
on
page
93.
Notes
92
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide