Applied Biosystems 7300 Installation And Maintenance Manual page 142

Real-time pcr system
Hide thumbs Also See for 7300:
Table of Contents

Advertisement

Appendix A
Moving the 7300/7500/7500 Fast System
®
Run a TaqMan
RNase P Instrument Verification
5.
plate (see
page
87).
• If the run passes, recalibrations are not
necessary.
• If the run fails, perform steps
through 9
to recalibrate the instrument.
Perform an ROI calibration (see
6.
Perform a Background calibration (see
7.
Perform an optical calibration if you are using a
7500 or 7500 Fast system (see
Perform a pure dye calibration (see
8.
Perform an instrument verification run (see
9.
page
87).
Notes
116
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
steps 6
page
35).
page
49).
page
59).
page
67).