Cleaning the Sample Wells
Wear powder-free
IMPORTANT!
gloves when you perform this
procedure.
Identify the contaminated wells of the sample
1.
block (see
"Troubleshooting" on page
Remove the plate and the tray holder.
2.
Close the tray. Apply pressure to the right side of
3.
the tray and at an angle.
Manually raise the block from the ROI Inspector
4.
window:
Click
(or select File New).
a.
In the New Document wizard,
b.
click
.
In the SDS software, select
c.
Instrument Calibrate.
In the warning dialog box, click
d.
lower the sample block.
The ROI Inspector dialog box opens.
In the ROI Inspector dialog box,
e.
click
Power off, then unplug the 7300/7500/7500 Fast
5.
system. Allow it to cool for 15 min.
Notes
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
64).
to
.
Appendix A
Decontaminating the Sample Block
Press
here
A
>00:15:00
109