Performing the Background Calibration
Load the plate in the instrument as described in
1.
"Loading the Plate" on page
If you cannot open the tray, the sample
Note:
block may be in its raised position, locking the
tray position. To lower the block, select
Instrument Calibrate, then exit the ROI
Inspector.
In the SDS software:
2.
Select the Instrument tab.
a.
Click
b.
The instrument begins the background
calibration run.
Before starting the run, the instrument
Note:
may pause (up to 10 minutes) to allow the heated
cover to reach the correct temperature.
Continue with
"Analyzing the Background
Calibration Data" on page
Notes
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Chapter 5 Performing the Background Calibration and Optical Calibration
40.
.
56.
Performing the Background Calibration