Performing the Verification Run

Load the plate in the instrument as described in
1.
"Loading the Plate" on page
When closing
Note:
the instrument tray,
apply pressure to the
right side of the tray
and at an angle.
In the plate document:
2.
Select the Instrument tab.
a.
Click
b.
The instrument begins the run, which lasts 1.5 to
2 hours using a 7300/7500 Real-Time PCR
system or <40 minutes using a 7500 Fast
Real-Time PCR system.
Before starting the run, the instrument
Note:
may pause (up to 10 minutes) to allow the heated
cover to reach the correct temperature.
When the run is complete, click
3.
Continue with
"Analyzing the TaqMan
Plate Data" on page
94.