In the ROI Inspector dialog box, right-click the
5.
black area of the window, then select Display
Properties.
In the Image Viewer Control Properties dialog box:
6.
Select Show Saturation.
a.
In the Sat Threshold field, enter 4000.
b.
Click
.
c.
In the ROI Inspector dialog box, set the lamp
7.
control:
Click
a.
Select Idle from the Lamp Control
b.
drop-down list (sets the lamp to lower
voltage for ROI calibration).
Continue with
"Performing the ROI Calibration" on
page
44.
Notes
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
Chapter 4 Performing the Regions of Interest (ROI) Calibration
.
Creating a Plate Document for the ROI Calibration
Select
a
b
c
a
b
43