Chapter 8 Powering Off the 7300/7500/7500 Fast System
Performing a Long-Term Shutdown
Performing a Long-Term Shutdown
Perform the long-term shutdown procedure if the
instrument will be inactive for more than 7 days.
Time Required
5 minutes
Materials Required
Plate, packaging (from
step 8 on page
Performing a Long-term Shutdown
Press the tray to open it.
1.
If the tray contains a plate, remove it.
2.
Load the packaging plate into the tray.
3.
If the shipping plate is not available,
Note:
substitute an unused reaction plate. During
storage, the instrument optics block rests on the
plate to protect the optics block.
Press the tray to move it into the instrument.
4.
Press the instrument power button.
5.
Power off the computer and monitor:
6.
Select
a.
In the Shut Down Windows dialog box (not
b.
shown), select
Power off the monitor.
c.
Notes
104
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
21)
Shut Down.
Shut Down.