Chapter 3 Installing the Software and Powering On the Instrument

Powering On the Instrument

Powering On the Instrument
Do not power on the instrument unless you have installed the SDS software. If you power on the
IMPORTANT!
instrument before you install the SDS software, the Windows operating system installs a generic driver for the
instrument, which does not allow the instrument to communicate with the SDS software.
Make sure that the computer Hibernate power setting is disabled. If the Hibernate setting is enabled,
IMPORTANT!
data collection stops when the computer goes into Hibernate mode. For more information, see
Settings and Power Options" on page
Powering On
Connect the Universal Serial Bus (USB) cable
1.
between the:
• USB connector on the back left of the
instrument
and
• Either USB port on the computer
Tower Computer USB Connection
Notes
28
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
137.
USB
"Setting the Display
Laptop Computer USB Connection
GR2374
7300/7500
electrical with laptop
USB