Appendix A

Replacing the Instrument Fuses

Replacing the Instrument Fuses

Time Required
Materials
Required
Replacing the
Fuses
Notes
122
Applied Biosystems 7300/7500/7500 Fast Real-Time PCR System Installation and Maintenance Guide
FIRE HAZARD. For continued protection against the risk of fire,
replace fuses only with listed and certified fuses of the same type and rating as those
currently in the instrument.
30 minutes
Fuses (2), 12.5A,
Flathead
250V, 5 x 20 mm
screwdriver
Turn off the instrument, then unplug it.
1.
Using a flat-head screwdriver, unscrew and remove the fuse holders from the
2.
instrument.
Powder-free
Safety
gloves
glasses